Michael Wallbank, 10/15/2018 12:57 PM


Front-end information

V1742 Digitizer

Post-Trigger

The V1742 has a register named POST_TRIGGER. This value is a percentage of the readout that is to be after the trigger. There is an implicit latency of around 40 ns, so in reality, the Post-Trigger percentage sets the percentage of the readout that will be after 40 ns past the trigger.

For an unknown reason, only certain values of POST_TRIGGER are accepted and must be found by trial and error.

Known working values

  • 0
  • 20
  • 45

MWPC Controller

Details:
  • Hardware address: 00.80.55.ee.00.01
  • IP address (programmed onto board!): 131.225.176.70
  • Name: ftbfwc03
  • CD system reference number: S22041
  • CD equipment ID number: 557912

Log in:

telnet ftbfwc03 <port>
Ports (according to here)
  • 5000 - daq apps
  • 5001 - emergency access
  • 5002 - telnet sessions

MWPC Configuration &