Attending: Susan, Micheal, Leon, Sandeep, Abbey, Mat, Mark,

DAQ Software- New IOC installed on NDOS.
- DDT work

Installation- Leon: Thursday removed 25 APD from Test stand
: Installed 30 new APDs one flaky APD.
: Swapped the bad APD on Saturday
: Removed those 30 and will install another 30 today
: ~20 APD available for NDOS install
Erik: Removed humidity sensor 1-5
Mat: MC decommissioning in planning

Holiday's: ?

Flahser: DCM 2-3 do see noise on the flasher channel. No news for the engineers.

OnMon: Try the OnMon again.
: Send Mark and Michael and dispatcher the machine and port for EventDispatcher.

Mon - Tue (13:41 12/10 - 10:11 12/11)
Run 14368, 21 subruns
No empty events, full subruns with expected number of events in each
FEB 2-2:60 shutoff in subrun 0, this is the same FEB which shutoff every chance it got during the previous week
28 other FEBs shutoff during subrun 17, sometime between 6:41 and 7:41. These were from DCMs 1-3 (3), 2-2 (11), 2-3 (1), 3-2 (9), and 3-3 (4).

Tue - Wed (20:28 12/11 - 10:45 12/12)
Run 14370, 15 subruns
No empty events, full subruns with expected number of events in each
One shutoff each in subruns 0, 5, and 7. These were all cooled and Silicon, Used Parylene, and New Parylene coated respectively.

Wed - Thu (14:07 12/12 - 13:33 12/13)
Run 14373, 24 subruns
Subrun 0 started out with a few empty events (15) before settling down.
Lost 2-1:46 and 3-1:22 in subrun 0, both were cooled and are silicon coated.
Bad things happened during subrun 19. Lost 73 FEBs and roughly 3000 active channels.
During subdun 20 lost 14 FEBs and another 18 during subrun 21.
FEB 2-2:10 which shutoff in subrun 19 and 2-2:42 which shutoff in subrun 20, are both reporting 4095 as a tecc.RVAL
In subrun 0 there were 8090 raw channels and 7455 active channels (the difference being the bad channel mask). In subrun 22 this had dropped to 4139 and 2241 respectively.

Thu - Fri (13:36 12/13 - 8:36 12/14)
Run 14376, 19 subruns
No empty events, full subruns with expected number of events in each
Lost 1-1:38 and 3-1:56 in subrun 0. These were uncoated not cooled, and Silicon coated cooled, respectively.
Lost 26 FEBs in subrun 1 and another 2 in subrun 2.

Fri - Sat (10:52 12/14 - 10:50 12/15)
Run 14378, 24 subruns
No empty events, full subruns with expected number of events in each
Lost 3-2:13 in subrun 4 which was New Parylene coated and cooled.
Great run!

Sat - Sun (10:50 12/15 - 10:48 12/16)
Run 14379, 24 subruns
No empty events, full subruns with expected number of events in each
Lost 113 FEBs in subrun 0, nearly half of all raw channels.
Lost 3 FEBs in subrun 1 and 1 more in subrun 13.
FEB 2-2:10 which shutoff in subrun 1 is still reporting a tecc.RVAL of 4095.
Not so good run.

Sunday (10:48 - 16:14)
Run 14380, 6 subruns
No empty events, full subruns with expected number of events in each
Lost 9 FEBs in subrun 1, 1 in subrun 2, 3 in subrun 3 and 2 in subrun 4.

First run in series was 14378 which started with 8210 raw channels, by the end of 14380 (54 hours later) less than half (4012) raw channels remain.
Someone (novadaquser) noticed that DCMs were missing on the EVD, stopped this run, and started a new one.

Sun - Mon (16:14 12/16 - 11:36 12/17)
Run 14381, 20 subruns
Subrun 0 started out with a few (68) empty events before settling down.
Lost 21 FEBs in subrun 0, 11 in subrun 16, 6 in subrun 17, and 34 in subrun 18.
Run started with 8180 raw channels, ended with 5906 raw channels.

Run 14382 was started immediately after 14381, I'm assuming people noticed missing DCMs in the EVD, but there was no entry to this affect.
This was a short run, only 6 minutes, so I can't speak to the effectiveness of this second restart to bring lost DCMs back into the fold.

Summary:

FEB 2-2:60 doesn't seem to be a problem past last Monday anymore.
Most runs have 0 empty events, 2 have very few at the beginning of subrun 0.
I'll highlight run 14378 which was a great run (Fri - Sat) only one FEB dropped out, zero empty events, and very stable in all other metrics. This goes downhill very quickly with the very next subrun.

14288 dcm name in hdb out of range

Run: 14364
From: 2012-12-07 22:13:43
To: 2012-12-08 22:11:37
Subruns: 24
FEB2-2:60 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-1:56 shutoff in subrun 0 is Silicon coated and was Cooled
FEB1-1:38 shutoff in subrun 1 is Un coated and was Not Cooled
FEB2-2:22 shutoff in subrun 12 is Used Parylene coated and was Cooled
FEB2-2:23 shutoff in subrun 12 is Used Parylene coated and was Cooled
FEB2-3:7 shutoff in subrun 12 is Silicon coated and was Cooled
FEB2-3:8 shutoff in subrun 12 is Silicon coated and was Cooled

Run: 14365
From: 2012-12-08 22:11:46
To: 2012-12-09 22:09:40
Subruns: 24
There were no shutoffs in this run

Run: 14366
From: 2012-12-09 22:09:48
To: 2012-12-10 10:48:27
Subruns: 13
FEB2-2:36 shutoff in subrun 8 is Used Parylene coated and was Cooled

Run: 14368
From: 2012-12-10 13:41:13
To: 2012-12-11 10:11:10
Subruns: 21
FEB2-2:60 shutoff in subrun 0 is Silicon coated and was Cooled
FEB1-3:14 shutoff in subrun 17 is New Parylene coated and was Cooled
FEB1-3:15 shutoff in subrun 17 is New Parylene coated and was Cooled
FEB1-3:32 shutoff in subrun 17 is New Parylene coated and was Cooled
FEB2-2:7 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB2-2:35 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB2-2:39 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB2-2:50 shutoff in subrun 17 is New Parylene coated and was Cooled
FEB2-2:55 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB2-2:56 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-2:57 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-2:58 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-2:59 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-2:61 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-2:62 shutoff in subrun 17 is Silicon coated and was Cooled
FEB2-3:46 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:1 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:2 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:3 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:7 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:9 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-2:22 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB3-2:25 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB3-2:28 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB3-2:29 shutoff in subrun 17 is Used Parylene coated and was Cooled
FEB3-3:1 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-3:4 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-3:7 shutoff in subrun 17 is Silicon coated and was Cooled
FEB3-3:39 shutoff in subrun 17 is Silicon coated and was Cooled

Run: 14370
From: 2012-12-11 20:28:49
To: 2012-12-12 10:45:31
Subruns: 15
FEB2-1:46 shutoff in subrun 0 is Silicon coated and was Cooled
FEB1-3:33 shutoff in subrun 5 is New Parylene coated and was Cooled
FEB3-2:27 shutoff in subrun 7 is Used Parylene coated and was Cooled

Run: 14373
From: 2012-12-12 14:07:13
To: 2012-12-13 13:33:04
Subruns: 24
FEB2-1:46 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-1:22 shutoff in subrun 0 is Silicon coated and was Cooled
FEB1-2:34 shutoff in subrun 19 is Silicon coated and was Cooled
FEB1-2:49 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB1-2:50 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB1-2:51 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB1-3:32 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB2-2:2 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:3 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:4 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:5 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:7 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:8 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:9 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:10 shutoff in subrun 19 is Used Parylene coated and was Error
FEB2-2:15 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-2:16 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB2-2:17 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:19 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:20 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:22 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:23 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:24 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:25 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:29 shutoff in subrun 19 is Used Parylene coated and was Not Cooled
FEB2-2:31 shutoff in subrun 19 is Used Parylene coated and was Not Cooled
FEB2-2:32 shutoff in subrun 19 is Silicon coated and was Not Cooled
FEB2-2:36 shutoff in subrun 19 is Used Parylene coated and was Not Cooled
FEB2-2:48 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:49 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:51 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:52 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:53 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:54 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:55 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB2-2:56 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-2:62 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:2 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:3 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:5 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:8 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:40 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:42 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:43 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-3:45 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-2:0 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-2:12 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-2:19 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:21 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:25 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:28 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:29 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:34 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB3-2:41 shutoff in subrun 19 is New Parylene coated and was Cooled
FEB3-2:49 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:50 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:52 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:53 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:54 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:55 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:56 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:57 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:59 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:60 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-2:61 shutoff in subrun 19 is Used Parylene coated and was Cooled
FEB3-3:2 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:10 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:13 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:34 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:37 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:38 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:39 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:40 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:41 shutoff in subrun 19 is Silicon coated and was Cooled
FEB3-3:44 shutoff in subrun 19 is Silicon coated and was Cooled
FEB2-2:42 shutoff in subrun 20 is Used Parylene coated and was Error
FEB2-2:43 shutoff in subrun 20 is Used Parylene coated and was Cooled
FEB2-3:15 shutoff in subrun 20 is Silicon coated and was Cooled
FEB2-3:38 shutoff in subrun 20 is Silicon coated and was Cooled
FEB3-2:13 shutoff in subrun 20 is New Parylene coated and was Cooled
FEB3-2:30 shutoff in subrun 20 is Used Parylene coated and was Cooled
FEB3-2:33 shutoff in subrun 20 is New Parylene coated and was Cooled
FEB3-2:35 shutoff in subrun 20 is New Parylene coated and was Cooled
FEB3-2:39 shutoff in subrun 20 is New Parylene coated and was Cooled
FEB3-2:40 shutoff in subrun 20 is New Parylene coated and was Cooled
FEB3-2:48 shutoff in subrun 20 is Used Parylene coated and was Cooled
FEB3-2:51 shutoff in subrun 20 is Used Parylene coated and was Cooled
FEB3-2:58 shutoff in subrun 20 is Used Parylene coated and was Cooled
FEB3-3:42 shutoff in subrun 20 is Silicon coated and was Cooled
FEB2-2:44 shutoff in subrun 21 is Used Parylene coated and was Cooled
FEB2-2:58 shutoff in subrun 21 is Silicon coated and was Cooled
FEB2-2:61 shutoff in subrun 21 is Silicon coated and was Cooled
FEB2-3:14 shutoff in subrun 21 is Silicon coated and was Cooled
FEB2-3:41 shutoff in subrun 21 is Silicon coated and was Cooled
FEB2-3:46 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-1:10 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-2:22 shutoff in subrun 21 is Used Parylene coated and was Cooled
FEB3-2:23 shutoff in subrun 21 is Used Parylene coated and was Cooled
FEB3-2:26 shutoff in subrun 21 is Used Parylene coated and was Cooled
FEB3-3:1 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:4 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:5 shutoff in subrun 21 is Silicon coated and was Not Cooled
FEB3-3:7 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:9 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:11 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:12 shutoff in subrun 21 is Silicon coated and was Cooled
FEB3-3:35 shutoff in subrun 21 is Silicon coated and was Cooled

Run: 14376
From: 2012-12-13 13:36:42
To: 2012-12-14 08:36:07
Subruns: 19
FEB1-1:38 shutoff in subrun 0 is Un coated and was Not Cooled
FEB3-1:56 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:49 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:52 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:53 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:56 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-2:57 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-2:58 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-3:40 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-3:41 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-3:42 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-3:43 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-2:0 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-2:28 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-2:35 shutoff in subrun 1 is New Parylene coated and was Cooled
FEB3-2:36 shutoff in subrun 1 is New Parylene coated and was Cooled
FEB3-2:50 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-2:54 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-2:55 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-2:56 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-2:58 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB3-3:1 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:9 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:14 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:34 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:35 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:36 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:38 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-3:2 shutoff in subrun 2 is Silicon coated and was Cooled
FEB3-3:44 shutoff in subrun 2 is Silicon coated and was Cooled

Run: 14378
From: 2012-12-14 10:52:41
To: 2012-12-15 10:50:41
Subruns: 24
FEB3-2:13 shutoff in subrun 4 is New Parylene coated and was Cooled

Run: 14379
From: 2012-12-15 10:50:41
To: 2012-12-16 10:48:41
Subruns: 24
FEB1-1:44 shutoff in subrun 0 is Un coated and was Not Cooled
FEB1-2:30 shutoff in subrun 0 is Silicon coated and was Cooled
FEB1-2:49 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB1-2:50 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB1-2:51 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB1-3:15 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB1-3:32 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB1-3:33 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB2-2:2 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:3 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:4 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:5 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:6 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:7 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:8 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:9 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:12 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:16 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB2-2:17 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:19 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:20 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:22 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:23 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:24 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:25 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:26 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:28 shutoff in subrun 0 is Used Parylene coated and was Not Cooled
FEB2-2:29 shutoff in subrun 0 is Used Parylene coated and was Not Cooled
FEB2-2:30 shutoff in subrun 0 is Used Parylene coated and was Not Cooled
FEB2-2:31 shutoff in subrun 0 is Used Parylene coated and was Not Cooled
FEB2-2:32 shutoff in subrun 0 is Silicon coated and was Not Cooled
FEB2-2:43 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:49 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:50 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB2-2:54 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:56 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:57 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:58 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:60 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:62 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:0 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:2 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:3 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:5 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:7 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:8 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:12 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:13 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:37 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:38 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:40 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:41 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:42 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:43 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:45 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:46 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-1:10 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-1:12 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-2:0 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-2:12 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-2:14 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:15 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:19 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:21 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:22 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:23 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:25 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:26 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:28 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:29 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:30 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:32 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:33 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:34 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:35 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:40 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:48 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:49 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:50 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:51 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:52 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:53 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:54 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:55 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:56 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:57 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:58 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:59 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:60 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:61 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-3:1 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:2 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:3 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:4 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:5 shutoff in subrun 0 is Silicon coated and was Not Cooled
FEB3-3:7 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:9 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:10 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:11 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:12 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:13 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:14 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:34 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:35 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:36 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:37 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:38 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:39 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:40 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:41 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:42 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:44 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:45 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-2:10 shutoff in subrun 1 is Used Parylene coated and was Error
FEB2-2:52 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:53 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:0 shutoff in subrun 13 is Used Parylene coated and was Cooled

Run: 14380
From: 2012-12-16 10:48:41
To: 2012-12-16 16:14:34
Subruns: 6
FEB2-2:15 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-2:34 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:39 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:46 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:48 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:51 shutoff in subrun 1 is Used Parylene coated and was Cooled
FEB2-2:59 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-3:34 shutoff in subrun 1 is Silicon coated and was Cooled
FEB3-1:56 shutoff in subrun 1 is Silicon coated and was Cooled
FEB2-2:44 shutoff in subrun 2 is Used Parylene coated and was Cooled
FEB3-2:5 shutoff in subrun 3 is Silicon coated and was Cooled
FEB3-2:7 shutoff in subrun 3 is Silicon coated and was Cooled
FEB3-2:9 shutoff in subrun 3 is Silicon coated and was Cooled
FEB2-3:14 shutoff in subrun 4 is Silicon coated and was Cooled
FEB2-3:33 shutoff in subrun 4 is Silicon coated and was Cooled

Run: 14381
From: 2012-12-16 16:14:42
To: 2012-12-17 06:14:55
Subruns: 14
FEB2-2:8 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:26 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB2-2:30 shutoff in subrun 0 is Used Parylene coated and was Not Cooled
FEB2-2:56 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:13 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:14 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:15 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:37 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:40 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:41 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:42 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:43 shutoff in subrun 0 is Silicon coated and was Cooled
FEB2-3:45 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-2:1 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-2:15 shutoff in subrun 0 is New Parylene coated and was Cooled
FEB3-2:19 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-2:21 shutoff in subrun 0 is Used Parylene coated and was Cooled
FEB3-3:1 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:5 shutoff in subrun 0 is Silicon coated and was Not Cooled
FEB3-3:9 shutoff in subrun 0 is Silicon coated and was Cooled
FEB3-3:11 shutoff in subrun 0 is Silicon coated and was Cooled