Using Synoptic for monitoring » History » Version 59

Flor de maria Blaszczyk, 02/27/2016 01:55 PM

1 1 Flor de maria Blaszczyk
{{toc}}
2 1 Flor de maria Blaszczyk
3 42 Jason St.john
*Synoptic plots are interactive:* 
4 42 Jason St.john
Zoom in by selecting a box around the area you're interested in.
5 42 Jason St.john
Zoom out by right-clicking on the plot.
6 42 Jason St.john
7 57 Flor de maria Blaszczyk
%{color:red}"Troubleshooting":https://cdcvs.fnal.gov/redmine/projects/lariat-online/wiki/Using_Synoptic_for_monitoring#Troubleshooting%
8 42 Jason St.john
9 1 Flor de maria Blaszczyk
h1. 1) How to start the Synoptic slow-control monitoring
10 39 Jason St.john
11 44 Flor de maria Blaszczyk
1) Open a console.
12 56 Flor de maria Blaszczyk
2) Run the *LArIATViewer.sh* script. This will get automatically a kerberos ticket with settings permissions and open the beamline overview.
13 44 Flor de maria Blaszczyk
14 56 Flor de maria Blaszczyk
*From ROC West:*
15 55 Flor de maria Blaszczyk
<pre>
16 55 Flor de maria Blaszczyk
cd /home/lariat/acnet
17 1 Flor de maria Blaszczyk
./LArIATViewer.sh
18 55 Flor de maria Blaszczyk
</pre>
19 55 Flor de maria Blaszczyk
20 56 Flor de maria Blaszczyk
21 56 Flor de maria Blaszczyk
*From the MCenter control room:*
22 54 Jason St.john
<pre>
23 54 Jason St.john
setup ftbf
24 54 Jason St.john
cd /home/nfs/ftbf_user/acnet
25 54 Jason St.john
./LArIATViewer.sh
26 54 Jason St.john
</pre>
27 44 Flor de maria Blaszczyk
28 44 Flor de maria Blaszczyk
29 44 Flor de maria Blaszczyk
Back up option: ACNET page *D99* launches Synoptic for Lariat.
30 1 Flor de maria Blaszczyk
31 43 Jason St.john
Detailed instructions (starting with launching ACNET): 
32 43 Jason St.john
If the Synoptic viewer is not opened already or if it has crashed, please follow the "steps 1 to 6 here":http://cdcvs.fnal.gov/redmine/projects/lardbt/wiki/How_to_use_the_online_monitoring#From-the-FTBF-MCenter-Control-Room-computers-any-desktop-computer
33 32 Flor de maria Blaszczyk
34 1 Flor de maria Blaszczyk
35 52 Flor de maria Blaszczyk
h1. 2) The displays main menu - Beamline overview
36 8 Flor de maria Blaszczyk
37 52 Flor de maria Blaszczyk
!BeamOverview.png!
38 1 Flor de maria Blaszczyk
39 52 Flor de maria Blaszczyk
This is the main menu. All displays/controls a shifter or expert needs can be accessed through this menu. 
40 1 Flor de maria Blaszczyk
41 52 Flor de maria Blaszczyk
* Click inside any of the frames (mouse-over until you get a small hand for a pointer) for each sub-system part to open the history plots for the corresponding devices. 
42 52 Flor de maria Blaszczyk
* The small gears on the top right corner inside the frames open the control panels when controls are available.
43 52 Flor de maria Blaszczyk
* Red boxes will appear to indicate a value is out of the allowed range.
44 52 Flor de maria Blaszczyk
* On top of the cryostat drawing, you might get alarm messages such as LAr dewar level too low, HV interlock or cryostat LAr level too low. Please pay attention to these.
45 1 Flor de maria Blaszczyk
46 52 Flor de maria Blaszczyk
* Click on the cryostat drawing to open the cryogenic system overview.
47 52 Flor de maria Blaszczyk
* Click on the TPC drawing to open the TPC overview.
48 1 Flor de maria Blaszczyk
49 1 Flor de maria Blaszczyk
*Beam related information*
50 1 Flor de maria Blaszczyk
* MC6CV indicates the opening a vertical collimator in the secondary beamline(see "Switchyard Layout":https://cdcvs.fnal.gov/redmine/attachments/download/23879/Switchyard-Current.pdf ). When the beam profiles do not look “good” (off-centered or too wide), MCR can change this value to improve the beam profile.
51 1 Flor de maria Blaszczyk
* MCENRG states the energy of the secondary beam at the LArIAT target.
52 1 Flor de maria Blaszczyk
53 1 Flor de maria Blaszczyk
*Tertiary magnets*
54 1 Flor de maria Blaszczyk
* MC7AT1 states the temperature of the coil of magnet 1. Alarm above 175º F.
55 1 Flor de maria Blaszczyk
* MC7AT2 states the temperature of the coil of magnet 2. Alarm above 175º F.
56 22 Flor de maria Blaszczyk
* +MC7AN+ is the current of the magnets . +Remember to report this number when starting a run+.
57 1 Flor de maria Blaszczyk
* MC7ANB is a probe inserted in our magnets to measure the polarity of the field. The actual value stated is not important as only the sign matters. +Double check that the polarity is the right one (positive or negative)+ required for the run and +remember to report its value in the elog when starting a run+.
58 2 Flor de maria Blaszczyk
59 20 Flor de maria Blaszczyk
*Scintillator counters*
60 2 Flor de maria Blaszczyk
* MC7SC1  states the number of counts for the scintillator counter before the LArIAT target.  Intensity of the secondary beam. 1E5 or more is pretty good.
61 8 Flor de maria Blaszczyk
* MC7SC2  states the number of counts for the scintillator counters after the last magnet.
62 1 Flor de maria Blaszczyk
* MC7SC3  states the number of counts for the scintillator counters after the Wire Chamber 4.
63 3 Flor de maria Blaszczyk
64 52 Flor de maria Blaszczyk
65 52 Flor de maria Blaszczyk
h1. 3) For shifters
66 52 Flor de maria Blaszczyk
67 52 Flor de maria Blaszczyk
Unless you need to change the settings on one of the power supplies / tertiary beam magnet, these are the displays you will need to look at.
68 52 Flor de maria Blaszczyk
69 52 Flor de maria Blaszczyk
h2. Beam related information
70 52 Flor de maria Blaszczyk
71 52 Flor de maria Blaszczyk
h3. MCenter Supercycle
72 52 Flor de maria Blaszczyk
73 52 Flor de maria Blaszczyk
To access this display, click on the "Beam info" frame on the beam-line overview.
74 52 Flor de maria Blaszczyk
75 52 Flor de maria Blaszczyk
!MCenterSupercycle.png!
76 52 Flor de maria Blaszczyk
77 52 Flor de maria Blaszczyk
On this display you can see at which stage of the beam cycle you are. The plots evolve continuously, wait for the spill (circled) to appear before stopping / starting a run (on the screenshot above you can see 2 spills). The full cycle is 1mn.
78 52 Flor de maria Blaszczyk
79 52 Flor de maria Blaszczyk
80 1 Flor de maria Blaszczyk
h3. Beam profiles
81 10 Flor de maria Blaszczyk
82 3 Flor de maria Blaszczyk
_Not available through Synoptic_ 
83 8 Flor de maria Blaszczyk
To get these for now you have to use ACNET. See "How to display the wire chambers positioned before the LArIAT target":https://redmine.fnal.gov/redmine/projects/lariat-online/wiki/Using_ACNET_for_monitoring#How-to-display-the-wire-chambers-positioned-before-the-LArIAT-target-along-the-secondary-beam-line .
84 8 Flor de maria Blaszczyk
85 28 Flor de maria Blaszczyk
h3. Beam-line detectors: Upstream and Downstream TOF, Halo and Punch-through, Aerogel Cerenkov
86 3 Flor de maria Blaszczyk
87 46 Sarah Lockwitz
!AGPMTsMonitor.png!
88 1 Flor de maria Blaszczyk
89 1 Flor de maria Blaszczyk
Because most of the PMTs for these systems are powered by the Zener voltage divider (aka COW) in the control room, controls and current monitoring are only available for the halo PMTs. These displays are similar, above is the one for the Aerogel Cerenkov.
90 3 Flor de maria Blaszczyk
91 1 Flor de maria Blaszczyk
_Note: If you cannot do settings on a page, it simply means no setting is possible. Pages where settings are possible are listed as "controls" in the main menu. Otherwise they are called "monitor"._ 
92 46 Sarah Lockwitz
93 52 Flor de maria Blaszczyk
h2. TPC Overview
94 12 Flor de maria Blaszczyk
95 52 Flor de maria Blaszczyk
To open this display, click on the TPC drawing in the beamline overview. Click inside the frames to get the history plots or on the gears to access the control panels.
96 9 Flor de maria Blaszczyk
97 52 Flor de maria Blaszczyk
!TPCOverview.png!
98 1 Flor de maria Blaszczyk
99 1 Flor de maria Blaszczyk
100 52 Flor de maria Blaszczyk
h2. Cryogenic system overview
101 21 Flor de maria Blaszczyk
102 52 Flor de maria Blaszczyk
To open this display, click on the cryostat drawing in the beamline overview. Click inside the frames to get the history plots.
103 21 Flor de maria Blaszczyk
104 52 Flor de maria Blaszczyk
!CryoOverview.png!
105 3 Flor de maria Blaszczyk
106 52 Flor de maria Blaszczyk
* An alarm will be triggered when the level inside the cryostat is below 2.5"Ar. 
107 52 Flor de maria Blaszczyk
* The level meter inside the LAr dewar does not cover the full range of the dewar. The maximum reading is 72"Ar which represent ~ 79% full. When there is a refill, the value should read 79% for a while, until the level reached the active range of the instrument.
108 52 Flor de maria Blaszczyk
* An alarm will be triggered when the LAr dewar level drops below 30%. 
109 52 Flor de maria Blaszczyk
110 59 Flor de maria Blaszczyk
_Note: Not all of the devices of the cryogenic system are shown, only a selection relevant for the LArIAT shifter._
111 1 Flor de maria Blaszczyk
112 1 Flor de maria Blaszczyk
h1. 4) For experts: Synoptic power supplies controls
113 1 Flor de maria Blaszczyk
114 12 Flor de maria Blaszczyk
Before you do anything... some useful information:
115 1 Flor de maria Blaszczyk
116 12 Flor de maria Blaszczyk
* *Settings are disabled by default.* This means you can zoom in / out on the plots or click on the links but cannot change any setting.
117 1 Flor de maria Blaszczyk
118 25 Flor de maria Blaszczyk
* *Closing a window restarts the plots, but the settings are NOT changed.* This means that if you close a window (on purpose or by mistake) or if the display crashes (it can happen from time to time) the power supply remains in whichever state it was before the closure. If it was on, it remains on at whatever voltage it was set. 
119 1 Flor de maria Blaszczyk
120 12 Flor de maria Blaszczyk
* *To power off a supply, please set the voltage to 0 first and then turn it off. Do not forget to turn it off!*
121 51 Flor de maria Blaszczyk
* *If a VME PS has to be turned off in an emergency situation (e.g. problem with TPC PMTs) - turn off the regulate function first, then +put the decreasing voltage values in the quick ramp window, go to 0 and then turn off!+*
122 1 Flor de maria Blaszczyk
123 12 Flor de maria Blaszczyk
!BiasControls.png! !GlassmanControls.png! !CryoPMTControls.png!
124 12 Flor de maria Blaszczyk
125 15 Flor de maria Blaszczyk
1) To enable settings, click on the button on the bottom right hand corner and select for how long you wish to enable the settings (Disable, 10mn, 1h or 8h). Do not forget to disable the settings once you are done! 
126 16 Flor de maria Blaszczyk
127 15 Flor de maria Blaszczyk
_There are pop-up confirmation windows that will prevent you from changing anything by mistake but it is still better to disable the settings once you have finished._ 
128 1 Flor de maria Blaszczyk
129 1 Flor de maria Blaszczyk
2) To turn on / off an power supply, click on the on / off button. A window will pop-up for confirmation. When the supply is on, the light should be green.
130 1 Flor de maria Blaszczyk
131 6 Flor de maria Blaszczyk
3) Turn on the regulate function when available (the light will turn green). This is to have an output voltage that matches the set voltage, otherwise there is a small offset. 
132 1 Flor de maria Blaszczyk
133 51 Flor de maria Blaszczyk
* *Ramps and regulate are not compatible*. Make sure regulate is off before using / setting a voltage on the ramp mode of a power supply.
134 51 Flor de maria Blaszczyk
* When regulate is off, there is an offset between the value you set and the actual readout. The lower the voltage you set (with respect to the range of the PS, ie 0-2kV or 0-500V), the larger the offset. 
135 51 Flor de maria Blaszczyk
* In particular, *for the induction plane bias, set a value of ~ 36-38 V to get 18V on the output*. Do not use regulate for this value, it will not work.
136 51 Flor de maria Blaszczyk
137 1 Flor de maria Blaszczyk
_In some cases this feature has been disabled because it was causing more trouble than help. You might have to play with the value you enter to get on the readout the value you want if you see a disagreement._
138 1 Flor de maria Blaszczyk
139 51 Flor de maria Blaszczyk
4) Click on the button in the green frame "Set voltage" (in volts) next to the power supply you would like to set a voltage to. A window will pop up, enter the value including the "-" sign and validate. If you enter a positive value and the supply is negative only, nothing will happen. You will have to re-enter a value. 
140 1 Flor de maria Blaszczyk
141 1 Flor de maria Blaszczyk
5) The selected supply readout voltage should indicate the new output voltage. Please wait until the reading stabilizes before changing to another value. This might take a few seconds.
142 1 Flor de maria Blaszczyk
143 51 Flor de maria Blaszczyk
6) Some controls (Cryo-PMTs) might include additional features such as slow ramp options. Please select the proper button according to the ramp time you wish to have. The Glassman has a ramp built in. 
144 1 Flor de maria Blaszczyk
145 12 Flor de maria Blaszczyk
Current can be readout in the bottom left hand side of the display. The top plot shows the voltage as a function of time, the bottom plot is the current as a function of time. To access the history plots, click on the "Last 24h" button.
146 51 Flor de maria Blaszczyk
147 51 Flor de maria Blaszczyk
_Note: The VME power supplies are not stable below a certain a value: for the 0-2kV ones (all PMTs, including cryo ones), do not set voltages below 200V; for the 0-500V (bias PS), do not set values below 36V (which will give you an 18V output)._
148 1 Flor de maria Blaszczyk
149 1 Flor de maria Blaszczyk
h1. 5) For experts: Controlling the power supplies with ACNET
150 1 Flor de maria Blaszczyk
151 1 Flor de maria Blaszczyk
*Please use this method ONLY if the Synoptic displays are not working even after restarting the Synoptic Viewer* (and contact F. Blaszczyk to report the issue). Once ACNET settings permissions are enabled, it is very easy to make mistakes: there is no confirmation window so you could cause a power supply to turn off or trip if you are not familiar with ACNET.
152 1 Flor de maria Blaszczyk
153 1 Flor de maria Blaszczyk
[[Using ACNET to power up the power supplies]]
154 1 Flor de maria Blaszczyk
155 1 Flor de maria Blaszczyk
h1. Troubleshooting
156 1 Flor de maria Blaszczyk
157 14 Flor de maria Blaszczyk
If Synoptic crashes or hangs up, do not panic! Only the application is affected, not the devices themselves. You only need to close the application and restart it. The non-historical plots will be reset.
158 1 Flor de maria Blaszczyk
159 30 Flor de maria Blaszczyk
*The most common reasons why the Synoptic viewer does not start, crashes or gives you random errors are:*
160 27 Flor de maria Blaszczyk
161 58 Flor de maria Blaszczyk
* You are not opening Synoptic correctly, i.e. you forgot to get a Kerberos ticket or you didn't open Synoptic with the script (or through the ACNET console) when working on a control room desktop computer. Please read the instructions carefully. 
162 1 Flor de maria Blaszczyk
163 58 Flor de maria Blaszczyk
* From your personal computer: you don't have the right Java environment. Make sure you have the "Sun / Oracle Java Runtime Environment (JRE)":http://www.oracle.com/technetwork/java/javase/downloads/index.html installed and up-to-date. OpenJRE / IcedTea does not seem to work well with Synoptic (unless you're lucky).
164 13 Flor de maria Blaszczyk
165 58 Flor de maria Blaszczyk
* From your personal computer: your Java security parameters are not configured correctly. Because of some recent changes on the Java platform, you might have to do some tweaks to the Java security parameters. If you get an error that says that the application cannot be launched because it is from an unknown source / signature, please do the following:
166 30 Flor de maria Blaszczyk
167 1 Flor de maria Blaszczyk
1) Open the Java control panel. How to access it varies according to the OS (Google is your friend). You might have to launch it with admin privileges.
168 1 Flor de maria Blaszczyk
169 1 Flor de maria Blaszczyk
2) Select the security tab.
170 1 Flor de maria Blaszczyk
171 1 Flor de maria Blaszczyk
3) Drop the security to medium level.
172 1 Flor de maria Blaszczyk
*Note: This might not be possible in recent java versions. If you only have the high and very high options, choose high.
173 1 Flor de maria Blaszczyk
174 1 Flor de maria Blaszczyk
4) Add the following links to the exceptions list:
175 1 Flor de maria Blaszczyk
http://synoptic.fnal.gov
176 1 Flor de maria Blaszczyk
http://www-bd.fnal.gov
177 1 Flor de maria Blaszczyk
178 31 Flor de maria Blaszczyk
5) Validate. You might have to restart the computer for these changes to take effect. _Note: Make sure you type "http" and not "https". The latter will not work._
179 1 Flor de maria Blaszczyk
180 30 Flor de maria Blaszczyk
181 30 Flor de maria Blaszczyk
*The most common reasons why you cannot do settings are:*
182 30 Flor de maria Blaszczyk
183 30 Flor de maria Blaszczyk
* You forgot to enable the settings or you went past the time the settings were enabled.
184 30 Flor de maria Blaszczyk
185 30 Flor de maria Blaszczyk
* There are no possible settings on the display (monitor only plots).
186 30 Flor de maria Blaszczyk
187 35 Flor de maria Blaszczyk
* You are not including the "-" sign in front of the value you are trying to set. All the negative VME power supplies require you to include the sign, while the Glassman does not.
188 34 Flor de maria Blaszczyk
189 36 Flor de maria Blaszczyk
* You are trying to set a value larger than the allowed limit. For security reasons or physical reasons (maximum PS voltage), some settings have been restricted to a particular range.
190 36 Flor de maria Blaszczyk
191 30 Flor de maria Blaszczyk
* Your Kerberos account has not been granted setting permissions. There is only a few people who have settings permissions on their individual accounts. If you really need to do settings, all desktop computers in the control room can do settings, provided you started Synoptic correctly.
192 30 Flor de maria Blaszczyk
193 30 Flor de maria Blaszczyk
* You are off-site. Even if you follow the procedure, and even if you're connected using the Fermilab VPN, you will not be able to enable settings. 
194 30 Flor de maria Blaszczyk
195 37 Flor de maria Blaszczyk
* We (as in LArIAT users) do not have settings permissions for that particular device. This shouldn't happen since all the devices used on Synoptic are LArIAT devices so we should have permissions for all of them. If it does happen, the device might be a recent one so we haven't been granted permissions yet. Please let Flor Blaszczyk and Charlie Briegel know. If you are doing settings on ACNET, make sure it is one of our devices and refrain from trying to change other devices (unless you are 100% sure you should). 
196 33 Flor de maria Blaszczyk
197 30 Flor de maria Blaszczyk
*Common web-viewer issues:*
198 30 Flor de maria Blaszczyk
199 30 Flor de maria Blaszczyk
* "I know the device is on, why does it show as if it were off?" Because of Fermilab firewalls or your respective institution firewalls, it might take couple minutes for the displays to show up. If you see little "stars" instead of values with a unit in the readout sections of the displays and that nothing shows on the time plots, it means you're not getting any information from the devices. Only way to go around or speed things up: use the Fermilab VPN.
200 1 Flor de maria Blaszczyk
201 1 Flor de maria Blaszczyk
h1. Contact
202 1 Flor de maria Blaszczyk
203 1 Flor de maria Blaszczyk
*LArIAT contact*: Flor de Maria Blaszczyk (Boston University), fblaszcz@bu.edu
204 1 Flor de maria Blaszczyk
Office phone #: 617-358-0989
205 1 Flor de maria Blaszczyk
Cell-phone # (also on the Control Room white board): 225-280-0536
206 1 Flor de maria Blaszczyk
Skype: flor.de.maria.blaszczyk
207 1 Flor de maria Blaszczyk
208 1 Flor de maria Blaszczyk
_Need a new display? Don't understand something? Please let me know! All suggestions and comments are welcome._
209 1 Flor de maria Blaszczyk
210 7 Flor de maria Blaszczyk
*Synoptic support*: synoptic-support@fnal.gov
211 19 Flor de maria Blaszczyk
212 53 Flor de maria Blaszczyk
*ACNET controls and VME power supplies emergencies*: Charlie Briegel, briegel@fnal.gov
213 19 Flor de maria Blaszczyk
214 53 Flor de maria Blaszczyk
*Synoptic only emergencies*: If no matter what you do, you cannot restart the OM, please contact *Linden Carmichael* (carmichael@fnal.gov) x5036 or 630-728-7059 (cell) or *Denise Finstrom* x4688