h1. 1) How to start the Synoptic slow-control monitoring
4 1 Flor de maria Blaszczyk
5 4 Flor de maria Blaszczyk
If the Synoptic viewer is not opened already or if it has crashed, please follow the steps 1 to 6 in the section "From the FTBF MCenter Control Room computers (ftbflx10 and ftbflx11)” in "Starting the Synoptic OM":https://cdcvs.fnal.gov/redmine/projects/lardbt/wiki/How_to_use_the_online_monitoring
6 1 Flor de maria Blaszczyk
7 5 Flor de maria Blaszczyk
*All plots are interactive: you can scroll in / out (hold Ctrl key while you scroll) with your mouse or select a "box"*
8 5 Flor de maria Blaszczyk
9 1 Flor de maria Blaszczyk
h1. 2) The displays main menu
10 1 Flor de maria Blaszczyk
11 8 Flor de maria Blaszczyk
!MainMenu.png!
12 1 Flor de maria Blaszczyk
13 5 Flor de maria Blaszczyk
This is the main menu, all displays/controls a shifter or expert needs can be accessed through this menu. Click on the menu to open the display / control panel you would like to look at. See the "For shifters" section for a full description of each display.
14 1 Flor de maria Blaszczyk
15 1 Flor de maria Blaszczyk
*Note: All the displays have a button that allows you to go back to the main menu, generally at the bottom right corner of the display.*
16 1 Flor de maria Blaszczyk
17 1 Flor de maria Blaszczyk
h1. 3) For shifters
18 1 Flor de maria Blaszczyk
19 5 Flor de maria Blaszczyk
Unless you need to change the settings on one of the power supplies / tertiary beam magnet, these are the displays you will need to look at.
20 1 Flor de maria Blaszczyk
21 1 Flor de maria Blaszczyk
h2. Beam related information
22 2 Flor de maria Blaszczyk
23 1 Flor de maria Blaszczyk
h3. MCenter Supercycle
24 2 Flor de maria Blaszczyk
25 8 Flor de maria Blaszczyk
26 1 Flor de maria Blaszczyk
On this display you can see at which stage of the beam cycle you are. The plots evolve continuously, wait for the spill (circled) to appear before stopping / starting a run. The full cycle is 1mn.
27 1 Flor de maria Blaszczyk
 
28 1 Flor de maria Blaszczyk
h3. MCenter info
29 2 Flor de maria Blaszczyk
30 3 Flor de maria Blaszczyk
You can access the tertiary beam magnets and chiller controls from this display.
31 2 Flor de maria Blaszczyk
32 8 Flor de maria Blaszczyk
!MCenterInfo.png!
33 1 Flor de maria Blaszczyk
34 3 Flor de maria Blaszczyk
*Beam related information*
35 8 Flor de maria Blaszczyk
* MC6CV indicates the opening of the vertical collimator XXXADD LOCATIONXXX. When the beam profiles do not look “good” (off-centered or too wide), MCR can change this value to improve the beam profile.
36 1 Flor de maria Blaszczyk
* MC7AN is the current of the magnets . Remember to report this number when starting a run.
37 1 Flor de maria Blaszczyk
* MCENRG states the energy of the secondary beam at the LArIAT target.
38 1 Flor de maria Blaszczyk
* MC7AT1 states the temperature of the coil of magnet 1. Alarm above 175º F.
39 1 Flor de maria Blaszczyk
* MC7AT2 states the temperature of the coil of magnet 2. Alarm above 175º F.
40 1 Flor de maria Blaszczyk
41 1 Flor de maria Blaszczyk
*Scintillator counters*
42 3 Flor de maria Blaszczyk
* MC7SC1  states the number of counts for the scintillator counter before the LArIAT target.  Intensity of the secondary beam. 1E5 or more is pretty good.
43 3 Flor de maria Blaszczyk
* MC7SC2 and MC7SC3 state the number of counts for the scintillator counters after the last magnet and after the Wire Chamber 4, respectively. 
44 3 Flor de maria Blaszczyk
* MC7SC5 states the number of counts for the coincidence of SC1, 2, 3.
45 8 Flor de maria Blaszczyk
* MC7SC6 used to be the "trigger": coincidence between wire chambers, TOF, halo and/or other beam elements.
46 3 Flor de maria Blaszczyk
47 1 Flor de maria Blaszczyk
h3. Beam profiles
48 1 Flor de maria Blaszczyk
49 10 Flor de maria Blaszczyk
_Not available through Synoptic_ 
50 10 Flor de maria Blaszczyk
To get these for now you have to use ACNET. See "How to display the wire chambers positioned before the LArIAT target":https://redmine.fnal.gov/redmine/projects/lariat-online/wiki/Using_ACNET_for_monitoring#section-3 . 
51 4 Flor de maria Blaszczyk
52 1 Flor de maria Blaszczyk
h3. Beam-line detectors: Upstream and Downstream TOF, Halo and Punch-through, Aerogel Cerenkov
53 3 Flor de maria Blaszczyk
54 10 Flor de maria Blaszczyk
!AGPMTsMonitor.png!
55 3 Flor de maria Blaszczyk
56 8 Flor de maria Blaszczyk
Because most of the PMTs for these systems are powered by the Zener voltage divider (aka COW) in the control room, controls and current monitoring are only available for the halo PMTs. These displays are similar, above is the one for the Aerogel Cerenkov.
57 8 Flor de maria Blaszczyk
58 3 Flor de maria Blaszczyk
59 1 Flor de maria Blaszczyk
h2. TPC 
60 1 Flor de maria Blaszczyk
61 1 Flor de maria Blaszczyk
h3. High-voltage summary
62 3 Flor de maria Blaszczyk
63 3 Flor de maria Blaszczyk
h3. Cryogenic system
64 9 Flor de maria Blaszczyk
65 3 Flor de maria Blaszczyk
The cryogenic system control is fully automated, these plots are for the shifter general interest.
66 1 Flor de maria Blaszczyk
67 1 Flor de maria Blaszczyk
*Overview* 
68 3 Flor de maria Blaszczyk
Shows an overall view of the current readings on different parts of the cryogenic system. For historical plots, see the individual displays.
69 9 Flor de maria Blaszczyk
70 9 Flor de maria Blaszczyk
_Note: Not all of the devices of the cryogenic system are shown, only a selection relevant for the LArIAT shifter._ 
71 3 Flor de maria Blaszczyk
72 3 Flor de maria Blaszczyk
*Cryostat info*
73 3 Flor de maria Blaszczyk
The plots in this display show the temperatures inside the cryostat, the pressures in different parts of the cryostat (the vacuum jacket and between the helico-flex seals in particular), and the liquid Argon level inside the inner cryostat for the last 24h.
74 3 Flor de maria Blaszczyk
75 3 Flor de maria Blaszczyk
*LAr and N duwars*
76 3 Flor de maria Blaszczyk
These plots show the evolution over the last 24h of the pressure and level of the outer LAr and N duwars.
77 3 Flor de maria Blaszczyk
78 3 Flor de maria Blaszczyk
*Filter and gas analyser*
79 5 Flor de maria Blaszczyk
For the gas analyser, you have to check the current configuration of the gas sampler. Please check the e-log for more information (?) .
80 1 Flor de maria Blaszczyk
81 1 Flor de maria Blaszczyk
h1. 4) For experts: Synoptic power supplies controls
82 1 Flor de maria Blaszczyk
83 1 Flor de maria Blaszczyk
*Settings are disabled by default.* This means you can zoom in / out on the plots or click on the links but cannot change any setting.
84 1 Flor de maria Blaszczyk
85 1 Flor de maria Blaszczyk
*Closing a window restarts the plots, but the settings are NOT changed.* This means that if you close a window (on purpose or by mistake) or if the display crashes (it can happen from time to time) the power supplies in whichever state it was before the closure. If it was on, it remains on at whatever voltage it was set. 
86 1 Flor de maria Blaszczyk
87 1 Flor de maria Blaszczyk
*To power off a supply, please set the voltage to 0 first and then turn it off. Do not forget to turn it off!*
88 1 Flor de maria Blaszczyk
89 1 Flor de maria Blaszczyk
xxx
90 1 Flor de maria Blaszczyk
91 1 Flor de maria Blaszczyk
1) To enable settings, click on the button on the bottom right hand corner and select for how long you wish to enable the settings.
92 1 Flor de maria Blaszczyk
93 1 Flor de maria Blaszczyk
2) To turn on / off an power supply, click on the on / off button. A window will pop-up for confirmation. When the supply is on, the light should be green.
94 1 Flor de maria Blaszczyk
95 6 Flor de maria Blaszczyk
3) Turn on the regulate function when available (the light will turn green). This is to have an output voltage that matches the set voltage, otherwise there is a small offset. 
96 6 Flor de maria Blaszczyk
97 7 Flor de maria Blaszczyk
_In some cases this feature has been disabled because it was causing more trouble than help. You might have to play with the value you enter to get on the readout the value you want._
98 1 Flor de maria Blaszczyk
99 1 Flor de maria Blaszczyk
4) Click on the button in the green frame "Set voltage" (in volts) next to the power supply you would like to set a voltage to. A window will pop up, enter the value including the "-" sign and validate. If you enter a positive value and the supply is negative only, nothing will happen. You will have to re-enter a value.
100 1 Flor de maria Blaszczyk
101 1 Flor de maria Blaszczyk
5) The selected supply readout voltage should indicate the new output voltage. Please wait until the reading stabilizes before changing to another value. This might take a few seconds.
102 1 Flor de maria Blaszczyk
103 1 Flor de maria Blaszczyk
6) Some controls (Glassman PS and Cryo-PMTs) might include additionnal features such as slow ramp options. 
104 1 Flor de maria Blaszczyk
105 1 Flor de maria Blaszczyk
Current can be readout in the bottom left hand side of the display. The top plot shows the voltage as a function of time, the bottom plot is the current as a function of time.
106 1 Flor de maria Blaszczyk
107 1 Flor de maria Blaszczyk
h1. 5) For experts: Controlling the power supplies with ACNET
108 1 Flor de maria Blaszczyk
109 1 Flor de maria Blaszczyk
*Please use this method ONLY if the Synoptic displays are not working even after restarting the Synoptic Viewer* (and contact F. Blaszczyk to report the issue). Once ACNET settings permissions are enabled, it is very easy to make mistakes: there is no confirmation window so you could cause a power supply to turn off or trip if you are not familiar with ACNET.
110 1 Flor de maria Blaszczyk
111 1 Flor de maria Blaszczyk
[[Using ACNET to power up the power supplies]]
112 1 Flor de maria Blaszczyk
113 1 Flor de maria Blaszczyk
h1. Troubleshooting
114 1 Flor de maria Blaszczyk
115 1 Flor de maria Blaszczyk
If Synoptic crashes or hangs up, do not panic! Only the application is affected, not the devices themselves. You only need to close the application and restart it. The non-historical plots will be reset.
116 1 Flor de maria Blaszczyk
117 1 Flor de maria Blaszczyk
Because of some recent changes on the Java platform, you might have to do some tweaks to the Java security parameters. If you get an error that says that the application cannot be launched because it is from an unknown source / signature, please do the following:
118 1 Flor de maria Blaszczyk
119 1 Flor de maria Blaszczyk
1) Open the Java control panel. How to access it varies according to the OS (Google is your friend). You might have to launch it with admin privileges.
120 1 Flor de maria Blaszczyk
121 1 Flor de maria Blaszczyk
2) Select the security tab.
122 1 Flor de maria Blaszczyk
123 1 Flor de maria Blaszczyk
3) Drop the security to medium level.
124 1 Flor de maria Blaszczyk
*Note: This might not be possible in recent java versions. If you only have the high and very high options, choose high.
125 1 Flor de maria Blaszczyk
126 1 Flor de maria Blaszczyk
4) Add the following links to the exceptions list:
127 1 Flor de maria Blaszczyk
http://synoptic.fnal.gov
128 1 Flor de maria Blaszczyk
http://www-bd.fnal.gov
129 1 Flor de maria Blaszczyk
130 1 Flor de maria Blaszczyk
5) Validate. You might have to restart the computer for this changes to take effect.
131 1 Flor de maria Blaszczyk
132 1 Flor de maria Blaszczyk
*Note: Make sure you type "http" and not "https". The latter will not work.*
133 1 Flor de maria Blaszczyk
134 1 Flor de maria Blaszczyk
h1. Contact
135 1 Flor de maria Blaszczyk
136 1 Flor de maria Blaszczyk
*LArIAT contact*: Flor de Maria Blaszczyk (Boston University), fblaszcz@bu.edu
137 1 Flor de maria Blaszczyk
Office phone #: 617-358-0989
138 1 Flor de maria Blaszczyk
Cell-phone # (also on the Control Room white board): 225-280-0536
139 1 Flor de maria Blaszczyk
Skype: flor.de.maria.blaszczyk
140 1 Flor de maria Blaszczyk
141 1 Flor de maria Blaszczyk
_Need a new display? Don't understand something? Please let me know! All suggestions and comments are welcome._
142 1 Flor de maria Blaszczyk
143 7 Flor de maria Blaszczyk
*Synoptic support*: synoptic-support@fnal.gov