The focus tests are to attempt to determine if our current "best focus" is truly the "best focus". Evidence suggests we are fairly close.
6 1 Kevin Reil
7 1 Kevin Reil
The test will consist of moving the focus set point to +25 microns for 10 exposures, -25 microns for 10 exposures and back to 0 microns for 10 exposures. This is done by issuing the following command via the Architect Console
8 1 Kevin Reil
9 4 Kevin Reil
h2. How
10 4 Kevin Reil
11 5 Kevin Reil
take 3 x 60 s exposures at with nominal settings (defaults) to get into focus/aligned.
12 6 Kevin Reil
<pre> FOCUS.AOS set set_point 0,0,25,0,0,0
13 6 Kevin Reil
 FOCUS.AOS set set_point 0,0,25,0,0,0 </pre>
14 1 Kevin Reil
take 10 x 60 s exposures
15 6 Kevin Reil
<pre> FOCUS.AOS set set_point 0,0,-25,0,0,0 
16 6 Kevin Reil
 FOCUS.AOS set set_point 0,0,25,0,0,0 </pre>
17 1 Kevin Reil
take 10 x 60 s exposures
18 6 Kevin Reil
<pre> FOCUS.AOS set set_point 0,0,0,0,0,0 
19 6 Kevin Reil
 FOCUS.AOS set set_point 0,0,25,0,0,0 </pre>
20 5 Kevin Reil
take 10 x 60 s exposures
21 1 Kevin Reil
22 4 Kevin Reil
h2. How (details)
23 4 Kevin Reil
24 1 Kevin Reil
Detailed instructions for issuing a FOCUS.AOS command can be found at [[https://cdcvs.fnal.gov/redmine/projects/decamops/wiki/Using_AOS_(Adaptive_Optics_System)]].
25 3 Kevin Reil
26 3 Kevin Reil
The command can be issued at any time during an exposure and will take effect on the following exposure. 
27 1 Kevin Reil
28 4 Kevin Reil
h2. When
29 1 Kevin Reil
30 4 Kevin Reil
h3. During second half of night (preferred)
31 4 Kevin Reil
32 1 Kevin Reil
The observer should pick a field and (good) filter and simply track on that field for the above 30 exposures. Please note expids in the elog.
33 3 Kevin Reil
34 4 Kevin Reil
h3. During obstac observations
35 3 Kevin Reil
36 3 Kevin Reil
The observer can easily enter these commands during obstac observations. Please note the relevant exids in the elog.